Chemical Research in Chinese Universities ›› 2001, Vol. 17 ›› Issue (3): 264-270.

• Articles • Previous Articles     Next Articles

Cobalt Determination by Means of Flow Injection On-line Sorbent Preconcentration-Flame Atomic Absorption Spectrometry with 1-Nitroso-2-naphthol as a Complexing Reagent

YIN Xue-feng, ALI Akbar, XU Guang-ming, LING Yun-yang, GAO Jian, YE Mei-ying   

  1. Department of Chemistry, Zhejiang University, Hangzhou 310027, P. R. China
  • Received:2000-04-20 Online:2001-08-24 Published:2011-08-04
  • Supported by:

    Supported by Zhejiang Natural Science Foundation of China.

Abstract: The suitability of 1-nitroso-2-naphthol(NN) as a complexing agent for on-line preconcen tration of cobalt eluted on the C18 microcolumn by means of the FI-FAAS system was tested.Various parameters affecting the complex formation and its elution were optimized.A 2.3 × 10-3 mol/L reagent solution and the aqueous sample solution acidified with 0.1 % (volume fraction) nitric acid were on-line mixed (6.4 mL/ min) on a reaction coil set at (65 ±1) ℃ and flowed through the microcolumn for 30 s.The pH of the mixed solution was adjusted to 3-4 with HNO3(1 mol/L) or NaOH(1 mol/L).The adsorbed complexes in the microcolumn were eluted into the nebulizer of FAAS in 10 s with ethanol acidified with 1 % HNO3(3.0 mL/min).A good precision(1.6 % for 100 μg/L Co(Ⅱ), n=10), a high enrichment factor 17.2, with detection limit (3σ) 3.2 μg/L, and sample throughput (90 h-1) were obtained.The method was applied to the certified reference materials(CRMs), NBS-362 and NBS-364, for the determination of cobalt and the results were in good agreement with the certified values.

Key words: Cobalt, FI on-line preconcentration, AAS, 1-Nitroso-2-naphthol