高等学校化学研究 ›› 1993, Vol. 9 ›› Issue (4): 303-308.
ZHANG Zheng-Qi, CHEN Sheng-Zong, LI Ning-Huai, ZHANG Hong
ZHANG Zheng-Qi, CHEN Sheng-Zong, LI Ning-Huai, ZHANG Hong
摘要: This paper covers the construction and behaviour of a mixed binder carbon paste electrode system appropriate for the cathodic stripping voltammetric quantitation of iso-niazid. The mixed binder consisted of glycerol and liquid paraffin. At the mixed binder carbon paste electrodes in a pH 3.0 buffer solution, isoniazid showed two sensitive cathodic stripping voltammetric wave at-0. 75 V (p1) and-0. 88 V (p2), respectively. The p2 can be used for the determination of trace amounts of isoniazid, the linear range of the peak current to the isoniazid concentration being from 5. 0×10-7 to 5.0×10-5 mol/L, and the limit of detection being 1. 0×10-7 mol/L with a relative standard deviation of 6. 0%(n=10). The proposed method was directly used to determine the drug in blood serum without the pretreatment of blood serum.